Development of a molecular platform for HTLV confirmatory diagnosis: importance of the internal amplification control
Retrovirology volume 8, Article number: A251 (2011)
Background
Brazil may harbor the largest absolute number of HTLV-1-infected individuals worldwide. The current HTLV diagnosis is based mainly on antibodies detection. However, these tests exhibit high proportion of indeterminate results. Several real-time PCR techniques have been developed for the detection of HTLV-1/2. However, up to day the major drawbacks of HTLV-1/2 molecular diagnosis are the lack of standard molecular tests and the absence of suitable internal amplification control (IAC). The aim of this study was to develop a multiplex qualitative real-time PCR for the simultaneous detection and discrimination of HTLV-1/2 and to design an IAC for reaction monitoring.
Methods
After multiple sequence alignments of the full genomes of HTLV-1/2 subtypes, a conserved tax region was chosen for the design of specific primers and probes. The IAC was generated after the annealing of synthetic nucleotide sequences and cloned into TOPO TA® vector. MT-2 and Gu cell lines were used as positive controls for HTLV-1 and HTLV-2, respectively.
Results
The developed multiplex real-time reaction detected both HTLV-1 and HTLV-2 (105 to 101 copies/reaction) at the presence of IAC in the same reaction. Analytical sensitivity was 1.2 copies/reaction for HTLV-1 and 19.1 copies/reaction for HTLV-2. The analytical sensitivity analysis was performed in singleplex format.
Conclusion
The detection of HTLV-1/2 and IAC by multiplex real-time PCR was efficient. The developed IAC is suitable for the molecular diagnosis and its presence ensures that the negative results are not due to failure in pre-PCR procedures.
Acknowledgements
Finnancial Support: CTC/INCTC, FAPESP, FUNDHERP, FINEP.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an open access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Rocha-Junior, M.C., Haddad, R., Rodrigues, E.S. et al. Development of a molecular platform for HTLV confirmatory diagnosis: importance of the internal amplification control. Retrovirology 8 (Suppl 1), A251 (2011). https://doi.org/10.1186/1742-4690-8-S1-A251
Published:
DOI: https://doi.org/10.1186/1742-4690-8-S1-A251
Keywords
- Multiple Sequence Alignment
- Antibody Detection
- Simultaneous Detection
- Molecular Diagnosis
- Molecular Test